DSpace
 

University of Jos Institutional Repository >
Medical Sciences >
Obstetrics and Gynaecology >

Please use this identifier to cite or link to this item: http://hdl.handle.net/123456789/1060

Title: Comparison of a New, Affordable Flow Cytometric Method and the Manual Magnetic Bead Technique for CD4 T-Lymphocyte Counting in a Northern Nigerian Setting
Authors: Imade, Godwin E.
Badung, Bitrus
Pam, Sunday
Agbaji, Oche
Egah, Daniel
Sagay, Atiene S.
Sankale, Jean-Louis
Kapiga, Saidi
Idoko, John
Kanki, Phyllis
Issue Date: Jan-2005
Publisher: Clinical and Diagnostic Laboratory Immunology
Series/Report no.: Vol.12;No.1; Pp 224-227
Abstract: We compared two techniques for CD4 T-lymphocyte counting: flow cytometry (Cyflow) and magnetic beads (Dynabead).Similar results with good correlation were obtained from the 40 adult blood samples counted(P=0.057, r=0.93). The Cyflow technique is more precise and cost-effective than the Dynabead method ($3 to $5 versus $12 to $22 per test, respectively), since as many as 200 samples can be measured per day.
URI: http://hdl.handle.net/123456789/1060
Appears in Collections:Obstetrics and Gynaecology

Files in This Item:

File Description SizeFormat
0216-04.pdf77.71 kBAdobe PDFView/Open
View Statistics

Items in DSpace are protected by copyright, with all rights reserved, unless otherwise indicated.

 

Valid XHTML 1.0! DSpace Software Copyright © 2002-2010  Duraspace - Feedback